How to Verify Directional Hearing Aids

By Interacoustics

 

 

Directionality Test using Real Ear Measurements

  1. From your patient management system, Noah or OtoAccess®, launch the Callisto™ Suite.
  2. Launch the REM440 module by clicking on the REM tab.
  3. Prepare the patient for Real Ear Measures.
  4. Calibrate the probe tube.
  5. If fitting an open hearing instrument, perform Calibrate for open fit.
  6. Ensure that the correct Ear is selected.
  7. To activate the Directional Test, go to: Menu | Setup | Show/hide protocols | US | Standard Directionality. Insert the checkmark inside the box in front of Standard Directionality. This will ensure that the Directionality test is available from the list of protocols drop down menu (f choosing US country defaults during the CallistoTM software installation, the Directionality test will be available by default)
  8. Choose the preferred Stimulus in the corresponding dropdown list.
  9. Change the input level using the slider if desired.
  10. Click START and follow the instructions on screen
  11. Place the client in front of the loudspeaker as shown on the picture. Click OK and front microphone curve will be recorded. New instructions will immediately appear. Turn the client around and click OK.

 

How to add the directionality test to an existing protocol

  1. Go to Menu | Setup | REM440 Setup.
  2. Choose a protocol to copy (ex: Adult Aided Response) from the drop down list.
  3. Click New to create a new protocol.
  4. Click Yes when the software asks if you want to copy the currently selected protocol (If selecting NO, you will have to create the protocol from scratch, which is beyond the scope of this document. For more info, please refer
    to the Instructions for Use document, Additional Info document or Quick Guide on Protocol Setup )
  5. Rename the Protocol.
  6. Add the Directionality test to the existing protocol.
  7. Edit the settings if desired by clicking on the Settings button
  8. Adjust the Settings if desired .

Presenter

Interacoustics

Available Training

AC440 Protocol Setup

There are many different demands and wishes regarding how an audiometer should both function and appear depending on the specific situation and national standards. One of the greatest advantages of the AC440 module is the flexibility, enabling you to tailor the system according to your specific preferences in the so-called “test protocols”. This article describes basic functionality on how to work with the AC440 Protocols.

Interacoustics2026-10-02

Acceptable Noise Level (ANL) Test

Learn how to perform the acceptable noise level (ANL) test with the CallistoTM device.

Interacoustics2026-10-02

Adult Hearing Aid Verification

Learn how to verifiy hearing aids in adults with the CallistoTM device, using real ear measurements (REM).

Interacoustics2026-10-02

Basic Audiometry

Learn how to do basic audiometry with the CallistoTM device.

Interacoustics2026-10-02

Binaural Real Ear Measurements

Learn how to perform binaural real ear measurements (REM) with the CallistoTM.

Interacoustics2026-10-02

BKB-SIN speech-in-noise test

In this quick guide, you will learn how to use the BKB-SIN speech-in-noise test provided with the Interacoustics CallistoTM audiometer. Read more.

Interacoustics2026-10-02

How to perform CROS and BiCROS fitting

Learn how to fit CROS and BiCROS hearing aids with the CallistoTM device.

Interacoustics2026-10-02

Customized Printout Setup

Customized Printout Setup

Interacoustics2026-10-02

How to perform FM transparency verification

In this CallistoTM quick guide, you'll learn how to perform FM transparency verification. Read more.

Interacoustics2026-10-02

How to perform frequency lowering verification

In this CallistoTM quick guide, you'll learn how to determine if you should apply frequency lowering and how to verify the frequency lowering feature.

Interacoustics2026-10-02

Verifying Ear Level FM Systems

Verifying Ear Level FM Systems

Interacoustics2026-10-02

Print Reports

Learn how to create customized print templates for the Callisto™, which can be linked to individual protocols for quick printing.

Interacoustics2026-10-02

QuickSIN Test

Learn how to perform the quick speech in noise (QuickSIN) test with the CallistoTM device.

Interacoustics2026-10-02

Real-Ear-to-Coupler Difference

Learn how to measure the real-ear-to-coupler difference (RECD) with the CallistoTM device.

Interacoustics2026-10-02

Speech in Noise (SIN) Test

Learn how to perform the speech in noise (SIN) test with the CallistoTM device.

Interacoustics2026-10-02

Speech in Quiet (SIQ) Test

Learn how to perform the speech in quiet (SIQ) test with the CallistoTM device.

Interacoustics2026-10-02