How to perform rotary chair testing with the Orion Reclining rotational chair

By Dr Liz Fuemmeler

Rotary chair testing is a mid-frequency test of vestibular function, specifically the horizontal vestibulo-ocular reflex (VOR), evaluating function from 0.01 Hz to 0.64 Hz. The assessment compares rotational chair (head speed) to eye movement speed and provides gain, phase, and symmetry data for evaluation.

Rotational chair testing offers several benefits in a diagnostic appointment for the dizzy patient. It is a standardized and repeatable stimulus that can be tolerated by the full age spectrum. It is especially beneficial in the evaluation of patients with suspected bilateral vestibular loss.

There are four main tests that can be completed utilizing the rotational chair:

  • Sinusoidal Harmonic Acceleration (SHA)
  • Step Velocity (Step testing)
  • VOR Suppression
  • Visual VOR

Protocol setup

In Protocol Management settings, you can select which of the tests you want to include in your test battery for your patient, as well as the order and details of the test.

Figure 1: Protocol Management screen. Tests can be included or rearranged for order here.

In each subtest, you can edit the test and choose specific test parameters. Below shows an example of the SHA assessment test parameters (Figure 2). You can check or select the frequencies you plan to test for the patient. The gear symbol next to each frequency allows for further detailed parameters, such as number of cycles (Figure 3).

In this example, the following settings are selected. Results displayed set to left eye, SHA velocity profile set to variable, and all frequencies from .01 Hz to .64 Hz are active. Finally, moving chair to zero and practice mode are disabled.
Figure 2: Test settings for SHA. Offers ability to include different frequencies.

In Figure 2, there are two additional options that can be enabled for the test protocol. The “Move To Zero When Test Is Complete” will physically recenter the chair to its home position when the cycle is completed. This may result in an additional half or full cycle for the patient to return to that home position of the chair.

The “Practice Mode” option will exclude the first half cycle of data collected into the data analysis. With practice mode off, it ensures that every piece of data is included in the analysis, in case of patient fatigue during testing.

The automatic default for both options is “off”.

Includes duration, which is automatically calculated, skip calibration, which can be on or off, fixation light, which can be enabled or disabled, save eye recording, which can be on or off, alongside adjustable sliders for frequency, peak velocity, peak acceleration, and number of cycles.
Figure 3: Detailed test settings for SHA testing, if you were to click on the gear next to test frequency 0.02 Hz.

Preparing for the test: Patient setup

VNG goggles

To accurately record the eye movements, videonystagmography (VNG) goggles should be on the patient’s face and the patient’s eyes should be calibrated. For the SHA, Step, and VOR Suppression tests, the cover of the goggles should also be on. For the Visual VOR test, the cover will be on, and a stimulus will be on the TV screen.

Patient position

The patient should be seated in the chair with head/goggles adhered to the headrest utilizing Velcro straps. This ensures that head movement is not providing any artifact for the response. The patient should be secured in the chair utilizing the seatbelt.

Chair position

The chair should be fully upright and not reclined. The chair should also be able to move freely and away from other people or objects. The software will also prompt you with the following safety checklist:

  • Is the patient’s lap belt securely fastened?
  • Is the patient’s head restrained to the head support?
  • Is the area around the chair (5 feet/1.5m) clear of obstructions?

How to perform the Sinusoidal Harmonic Acceleration (SHA) test

The SHA test evaluates VOR gain, phase, and symmetry when the patient is rotated at a variety of speeds (0.01 Hz – 0.64 Hz) in alternating directions.

Example patient instructions: “The chair is going to rotate, please keep your eyes open wide. I am going to ask you some questions to keep your mind active during the test.”

After choosing the test protocol (frequencies included in the evaluation) and setting up the patient correctly in the chair, the test is ready to be started. While the patient is being rotated in the chair, their eyes need to be open, and the clinician will need to provide verbal tasking exercises to cognitively engage the brain and ensure the eye response cannot be suppressed.

Tasking examples are provided in Table 1.

Type of task Examples
ABC task Girls/boys names A-Z, cities/states, celebrities, fruit or vegetables…
Naming task

Name 3 things that are green, blue, soft, squishy…

Name 3 things you can wear… on your wrist, head, feet…

Name 3 things you would find in the… product section, living room, kitchen, train station…

Name 3 streets you drive on, on a daily basis

Name 3 types of cars

Counting task Count backward from 100 in steps of 7
Take me on a tour Take me on a tour of your house
Talk me through Talk me through making your favorite recipe

Table 1: Tasking examples to use in rotational chair assessment.

Understanding measurement parameters in the SHA test

Colors

Following completed SHA testing, the following screen will summarize the data collected (Figure 4).

Figure 4: SHA test summary screen.

Data is displayed for one eye at a time, which can be toggled at the top of the screen. The example above shows left eye data, which is why the data points and eye tracings are blue. Right eye data are shown in red.

Graphs

Gain, phase, and symmetry are calculated based on the data collected as displayed in Figure 4. If a data point is in the white region of the graph, it is within normal limits. If a data point is in the gray region, it is outside of normal limits. Normative data limits change based on age and age-based data can be found in “System Default Settings” > “Threshold Levels”.

Gain is the comparison of peak eye velocity to peak chair velocity at each frequency. Remember, this is the measurement of horizontal VOR gain. Phase is the measurement of the time delay (in degrees) between eye and chair velocity. Symmetry compares eye velocity measured in rightward chair versus leftward chair rotations.

The raw eye data is displayed for each frequency (eye position graph), and a trend line is established on an eye velocity graph. Lastly, the spectral purity percentage is displayed at the bottom right of the eye velocity graph, which is a general measure of test and data validity.

Editing SHA data

In the side panel, there is an “Edit Tools” option to refine data to exclude outliers or artifactual data, if needed. If you click the “Edit Tools” button, it opens several options to refine SHA data (Figure 5-10).

Figure 5: Enable section.

Figure 6: Disable section.

Figure 7: Overall gain increase.

Figure 8: Overall gain decrease.

Figure 9: Individual cycle gain adjustment, increase or decrease.

Figure 10: Shift phase.

How to perform the Step Velocity test

The Velocity Step Test measures the patient’s VOR by rotating the patient at an acceleration impulse to a fixed chair velocity with vision denied. Step tests are performed at 50 degrees per second and 180 degrees per second by default. The time constant, response gain, and time constant asymmetry are then measured.

By measuring the vestibular system’s time constant, both the peripheral vestibular response to the rotational stimulus and the central velocity storage mechanism can be evaluated, making this a useful test for aiding in the diagnosis of a variety of vestibular disorders.

Example patient instructions: “The chair is going to rotate, please keep your eyes open wide. I am going to ask you some questions to keep your mind active during the test.”

Refer to the tasks in Table 1 to help keep the patient engaged during testing.

Understanding measurement parameters in Step testing

Colors

Following completed Step testing, the following screen will summarize the data collected (Figure 11).

Figure 11: Step Velocity summary screen.

Data is displayed for one eye at a time, which can be toggled at the top of the screen. The example above shows left eye data, which is why the data points and eye tracings are blue. Right eye data are shown in red.

Graphs

Gain, time constant, and time constant symmetry are displayed following the test.

Gain is the comparison of peak eye velocity to peak chair velocity at each frequency. It is displayed for both per-rotary (after initiation of the chair movement) and post-rotary (after the chair stops), which is indicated by different symbols on the graph.

Time constant is the amount of time for the slow-phase velocity of the nystagmus to decrease by 63% (or 37% of its peak value) after the head stops rotating.

Time constant asymmetry compares the time constant between right and leftward movements.

How to perform the VOR Suppression Test

The VOR suppression test evaluates the ability of the brain to suppress, or overcome, the VOR reflex during head/whole-body movement of the chair. The patient is seated in the rotational chair with the goggles and cover on. They will see the fixation light inside the goggles while the chair rotates.

The SHA test must be completed prior to the VOR suppression test, as the outcome of this test compares the VOR gain without fixation versus the VOR gain with fixation. Specific frequencies can be chosen for this evaluation and should be a frequency that was tested in SHA. Because the patient has a task to keep their eyes on the fixation light, there is no need to provide verbal tasks during testing.

Example patient instructions: “You are going to see a light inside your goggles, stare at that light while the chair moves.”

Understanding measurement parameters in VOR Suppression testing

Colors

Following completed VOR suppression testing, the following screen will summarize the data collected (Figure 12).

Figure 12: VOR Suppression summary screen.

Data is displayed for one eye at a time, which can be toggled at the top of the screen. The example below shows left eye data, which is why the data points and eye tracings are blue. Right eye data are shown in red.

Graphs

Gain and gain reduction graphs are shown following the test.

The gain graph includes SHA VOR gain (shown with a blue circle in Figure 12) and VOR Suppression gain (shown with a blue triangle). VOR suppression gain should be low and eye tracing graphs should show little to no nystagmus (looks like a flat line) in normal cases.

The gain reduction graph is shown on the summary screen. Normal VOR cancellation should be in the white area of the graph. Abnormal VOR suppression data would show in the gray shaded region.

How to perform the Visual VOR test

The visual VOR test requires the patient to be moved in a pendular motion, as in sinusoidal harmonic acceleration with the addition of a visual stimulus on screen.

The fixation cover is removed from the goggles for this test and the patient is required to focus on a stationary target on the oculomotor screen during chair rotation from right to left and vice versa. You can perform the test at any SHA frequency, but it is most performed at 0.32 Hz.

Example patient instructions: “You will feel yourself moving back and forth slowly in the chair. During the rotation, you will see a target appear on the screen in front of you. Please keep your eyes focused on the target during the entire test.”

Understanding measurement parameters in the Visual VOR test

Colors

Following completed VOR suppression testing, data is displayed for one eye at a time, which can be toggled at the top of the screen. Left eye data are blue. Right eye data are red.

Graphs

Gain, phase, and symmetry are displayed following the test. Gain (%) is the ratio of maximum slow phase eye velocity to maximum head (chair) velocity. Phase (°) displays the timing relationship between the motion stimulus and the ocular response to that stimulus. Symmetry compares rightward to leftward movements.

Learn more about rotary chair testing

For additional resources and information on our rotational chair tests, we recommend reading the following blogs on our website:

References

Jacobson, G. P., & Shepard, N. T. (2015). Balance function assessment and management (2nd ed.). Plural Publishing.

Presenter

Dr Liz Fuemmeler

Dr Liz Fuemmeler

Dr. Liz Fuemmeler is the Balance Clinical Product Manager with Interacoustics and Vestibular Program Director at Professional Hearing Center in Kansas City, MO. She graduated with her Doctorate of Audiology in 2019 from Purdue University and received specialty training in vestibular and balance disorders at Boys Town National Research Hospital and the Mayo Clinic. She is actively involved in vestibular and concussion research and regularly lectures for local, national, and international conferences. Outside of her role with Interacoustics, she co-hosts a monthly podcast called "A Dose of Dizzy'' that reviews current vestibular protocols and research. She also is the past-president of the Missouri Academy of Audiology and volunteers with the American Academy of Audiology.

Available Training

No articles match the selected filters.

Looking for clinical training?

Are you looking for more advanced and clinical training?