Daily System Check

By Interacoustics

Why should I perform a daily system check/probe integrity test?

It is recommended to perform a daily check of your OAE equipment to ensure that it is in good working order, before testing on patients. As the probe and probe tip often come into contact with wax or other debris in the ear canal, a common fault affecting testing can be a blocked or partially obstructed probe tip. Running a probe integrity test and a real-ear check allows for any probe faults or system distortions that can masquerade as biological responses to be detected. A daily check ensures that you can be certain that the results obtained throughout the day are valid.

Probe integrity test

The probe integrity test ensures artifact responses (system distortions) are not being generated by the probe or hardware. The probe tip should be inspected for wax or debris prior to conducting the test. Testing should always be conducted in a quiet environment.

Test procedure

  1. Insert the probe into the provided test cavity or ear simulator. It's important to use the correct-sized cavity for valid test results. Depending on the test cavity, an ear tip may need to be attached to the probe tip first.

    For DPOAE a 0.2cc cavity is recommended.
    For TEOAE, a 0.5cc cavity is recommended.

    The infant ear simulator is specifically designed for assessing the probe with SeraTM

  2. Select an OAE test protocol. As system distortion is dependent on stimulus output level, select a protocol that reflects what is used most often in clinical practice.
  3. Start the test and allow it to run until the test stops automatically. Do not stop the test manually

Test results

If the probe is functioning correctly, none of the frequency bands (TEOAE) or points (DPOAE) should have a checkmark, i.e., no artifacts/OAEs should be detected above the noise floor.

If an error message appears during testing or if one or more of the OAE bands or points has a checkmark (meaning detected), the probe integrity test has failed. This could indicate that:

  1. There is wax or debris in the probe tip and cleaning is required.
  2. The probe was not placed in the test cavity or ear simulator correctly, or,
  3. The probe's calibration needs to be checked.

Check and clean the probe tip and retry the test. If the probe test fails a second time, the probe should not be used to test on patients. Contact your local service technician for assistance.

Real-Ear Check

This test can be done by placing the probe in one's own ear and running a commonly used test protocol.

If OAE results do not match the tester's expected OAE result, this could be an indication that:

  1. The probe is not connected correctly to the device.
  2. The ear tip is not attached correctly to the probe tip.
  3. There is wax or debris in the probe tip and cleaning is
    required.
  4. The environment is too noisy for testing.
  5. The probe was not placed correctly in the ear canal.
  6. The probe's calibration needs to be checked.

If results from the real-ear test do not match the expected result after checking items 1 to 5 above, the probe should not be used to test on patients. Contact your local service technician for assistance

Presenter

Interacoustics

Available Training

Creating an ABRIS Test Protocol

Learn how to use the Titan's automated ABR protocol.

Interacoustics2026-10-02

Acoustic Reflex Testing

Learn how to perform acoustic reflex testing with the Titan tympanometer.

Interacoustics2026-10-02

Best Practices for OAE Testing

Learn how to get the most out of your Titan for OAE testing.

Interacoustics2026-10-02

How to Identify a Blocked Probe Fit

Learn how to identify a blocked probe fit with the Titan tympanometer.

Interacoustics2026-10-02

How to Change the Probe Tip

Learn how to change the Titan's probe tip.

Interacoustics2026-10-02

How to Clean the Probe Tip

Learn how to clean the Titan's probe tip.

Interacoustics2026-10-02

Creating a DPOAE Pass/Refer Test Protocol

Learn how to create a DPOAE pass/refer test protocol with the Titan.

Interacoustics2026-10-02

Creating a TEOAE Pass/Refer Test Protocol

Learn how to create a TEOAE pass/refer test protocol with the Titan.

Interacoustics2026-10-02

Electrode Tips & Tricks

Learn how to get the most out of your electrodes for testing with the Titan.

Interacoustics2026-10-02

How to Identify Fluid in the Middle Ear

Learn how to identify fluid in the middle ear with the Titan tympanometer.

Interacoustics2026-10-02

How to clean the Titan probe tube

In this video, you will learn how to clean the Titan's probe tube.

Interacoustics2026-10-02

How to print without a database

Click here to learn about the three options you have for printing without a database with the Titan handheld tympanometer.

Interacoustics2026-10-02

How to Identify a Loose Probe Fit

Learn how to identify a loose probe fit with the Titan tympanometer.

Interacoustics2026-10-02

Mastoid Electrode Montage

Learn how to prepare the patient's skin and apply mastoid electrodes for automated ABR testing with the Titan.

Interacoustics2026-10-02

How to Measure eSRT

Learn how to measure electrically evoked stapedial reflex thresholds (eSRTs) with the Titan.

Interacoustics2026-10-02

Nape Electrode Montage

Learn how to perform nape electrode montage.

Interacoustics2026-10-02