BKB-SIN speech-in-noise test

By Interacoustics

In this quick guide, you will learn how to use the BKB-SIN speech-in-noise test provided with Interacoustics audiometers.

Getting started

  • Activate your speech-in-noise license key to get access to the test through your audiometry module.
  • Present the test via earphones or in a sound field with the attenuator dial set to 70 dB HL.
  • For subjects with 3F PTA hearing losses greater than 45 dB HL, set the attenuator dial to a level that is "Loud, But OK."
  • When testing in a sound field, have the patient speak into the talkback microphone so that responses are clearly audible to the tester.
  • Instruct the patient to repeat the sentences spoken by the main talker and to ignore the background talkers.
  • A verbal "ready" cue precedes each sentence.

Suggested presentation levels for standard SNR loss testing

Patient group Set dial to
Adult with normal hearing or mild loss 70 dB HL*
Adult with 3F PTA > 45 dB HL Loud, But OK
Child with normal hearing or mild loss 50 to 70 dB HL
Child with 3F PTA > 45 dB HL Loud, But OK
Patient aided with cochlear implant 50 to 70 dB HL

*At this level, the frequent peaks of the talker are at approximately 83 dB SPL for insert earphones, 85 dB SPL at 0 degrees azimuth in sound field, and 90 dB SPL with TOH earphones (ANSI S3.6 Specification for Audiometers, 1996).

Categories of loudness

Category Loudness
7 Uncomfortably Loud
6 Loud, But OK
5 Comfortable, But Slightly Loud
4 Comfortable
3 Comfortable, But Slightly Soft
2 Soft
1 Very Soft

Scoring

  • Both lists of a List Pair (A & B) must be administered for valid scoring.
  • Present List A.
  • Give one point for each underlined word repeated correctly.
  • Add the number of correct words.
  • Subtract the total correct from 23.5 (this is the SNR-50).

Repeat above steps for List B.

  • Average the scores for A & B to obtain the List Pair Score.

Note: If administering more than one List Pair, average all List Pair scores. Compare score to test norms to calculate SNR Loss (see reverse).

Scoring example

BKB-SIN test norms in adults (List Pairs 1-8)

Mean SNR-50 Standard deviation
Normal hearing adults Minus 2.5 0.8
Adults with cochlear implants Compare to normal-hearing adult value to determine SNR loss 1.6

BKB-SIN test norms in children (List Pairs 1-8)

Mean SNR-50 Standard deviation
Children aged 5-6 3.5 2.0
Children aged 7-10 0.8 1.2
Children aged 11-14 Minus 0.9 1.1

Calculate SNR loss

Subtract the test norm from the SNR-50 score. For example, if the SNR-50 is 9 dB, then the SNR loss would be calculated as follows:

Patient group Calculation SNR loss
Adult 9 minus (minus 2.5) 11.5 dB
Children aged 5-6 9 minus 3.5 5.5 dB
Children aged 7-10 9 minus 0.8 8.2 dB
Children aged 11-14 9 minus (minus 0.9) 9.9 dB

Interpretation (adults)

SNR loss Degree of SNR loss Expected improvement with directional microphones
0-3 dB Normal/near normal May hear better than normal in noise
3-7 dB Mild SNR loss May hear almost as well as normal in noise
7-15 dB Moderate SNR loss Directional microphones would help
Consider array microphone
>15 dB Severe SNR loss Maximum SNR improvement is needed
Consider FM system

Interpretation (children)

The effects of SNR loss are more pronounced for children than adults. Adults possess linguistic proficiency and world knowledge that facilitates speech perception, whereas children are in the process of acquiring linguistic and world knowledge through audition. Interpreting test results for children should be done a case-by-case basis.

Factors such as the student's speech, language and academic skills, as well as the learning environment (class size, acoustics properties of the room, and teaching style) should be considered. For example, a 5 dB SNR loss may have less impact on a child who has strong communicative and academic skills than one who is struggling in those areas.

Similarly, classroom environments that are instructionally and acoustically hospitable can sometimes reduce the adverse impact of the SNR loss. While those that are not may increase the impact of the SNR loss.

Learn more

To learn more about BKB-SIN testing, please read the BKB-SIN manual.

Presenter

Interacoustics

Available Training

Acceptable Noise Level (ANL) Test

Learn how to perform the acceptable noise level (ANL) test with the Affinity Compact.

Interacoustics2026-10-02

AC440 Protocol Setup

Learn how to set up protocols in the Affinity Compact's audiometry module.

Interacoustics2026-10-02

Adult Hearing Aid Verification

Learn how to verifiy hearing aids in adults with the Affinity Compact, using real ear measurements (REM).

Interacoustics2026-10-02

Affinity Compact: A product introduction

This video provides an introduction to and overview of the Affinity Compact.

Interacoustics2026-10-02

Affinity Compact: Features and functions

This video provides an overview of the Affinity Compact's features and functions.

Dennis Mistry2026-10-02

Ambient Noise

Worried about noise in your audiometry recordings? Affinity Compact has an Ambient Noise feature, which allows you to monitor for noise in the test environment.

Interacoustics2026-10-02

IA-AMTAS

Learn about the IA-AMTAS feature in Affinity Compact for automated audiometry testing.

Interacoustics2026-10-02

Basic Audiometry

Learn how to do basic audiometry with the Affinity Compact.

Interacoustics2026-10-02

How to perform CROS and BiCROS fitting

Learn how to fit CROS and BiCROS hearing aids with the Affinity Compact.

Interacoustics2026-10-02

Customized Printout Setup

Learn how to customize printouts with the Affinity Compact.

Interacoustics2026-10-02

Delta Values

To make hearing aid fitting easier, the Delta Values feature calculates the difference between your measurement curve and your target curve. Learn more here.

Interacoustics2026-10-02

How to perform FM transparency verification

In this Affinity Compact quick guide, you'll learn how to perform FM transparency verification. Read more.

Interacoustics2026-10-02

Free Field Adjustment

Learn about Affinity Compact's free field adjustment feature, which ensures your free field speaker is properly calibrated, no matter the sound environment.

Interacoustics2026-10-02

How to perform frequency lowering verification

In this Affinity Compact quick guide, you'll learn how to determine if you should apply frequency lowering and how to verify the frequency lowering feature.

Interacoustics2026-10-02

Hearing Aid Transition (HAT)

Learn about the hearing aid transition test, which allows you to copy tuning preferences from hearing aids which you do not have previous sessions stored for.

Interacoustics2026-10-02

HIT440 Protocol Setup

Learn how to set up protocols in the Affinity Compact's hearing instrument testing module.

Interacoustics2026-10-02
Affinity Compact | BKB-SIN speech-in-noise test | Interacoustics