How to perform audiometric masking with Equinox Evo

By Interacoustics

Required equipment

  • Equinox Evo
  • A patient response button
  • A calibrated bone conductor (B81)
  • Licensed AC440 Audiometry module
  • Calibrated headphones or insert phones

Test procedure

Perform air conduction and bone conduction audiometry unmasked for both ears.

In case there is an air-bone gap of 15 dB or more, the bone conduction threshold must be reassessed while applying masking to the non-test ear.

If the difference between the air conduction threshold of the worse ear and the bone conduction threshold of the good ear exceeds an interaural attenuation of 40 dB (50 dB if using insert phones), masking will be needed for the air conduction audiometry of the worse ear.

Select NB on channel 2.

Select the ear to be masked (right or left) and choose the appropriate masking transducer (headphones or insert phones). This will activate the masking noise.

Proceed to do a masked threshold search and press Store once a correct threshold has been obtained. The symbol will appear in the audiogram as masked.

There are many ways of applying clinical masking. Which one to use is your decision. Regardless of the masking method, channel 2 is used to occupy the better ear.

When storing a threshold while masking, the final masking level is stored in the masking table under the ear that is being tested. The term 'Effective masking' in this situation refers to the fact that the narrowband noise level was loud enough to effectively mask a pure tone of the indicated level heard by the masked ear.

Auto masking

Interacoustics auto masking is available to ease the effort required to mask with correct masking levels. When auto masking is enabled, channel 2 is controlled by the system and is set to the appropriate intensity level.

Auto masking is activated by selecting the icon that shows the mask with a letter A. There will be a green, amber, or purple indicator at the level of channel 2 indicating the following (Table 1).

Indicator Meaning
Green Masking is correctly applied.
Amber Masking should be louder, and extended range needs to be activated.
Purple Masking is needed but not possible.

Table 1: Auto masking indicators.

Masking help

Interacoustics masking help is available to make it easier to decide on a safe and correct masking intensity. When masking help is activated, a status light on channel 2 indicates if masking is applied correctly.

Masking help is activated by selecting the icon that shows the mask without a letter A. There will be a grey, green, amber, or purple indicator at the level of channel 2 indicating the following (Table 2).

Indicator Meaning
Grey Masking help is not active.
Green Masking is correctly applied. May also provide a suggested masking level.
Amber Masking should be increased or decreased to suggested level.
Purple Masking is needed but not possible.

Table 2: Masking help indicators.

Presenter

Interacoustics

Available Training

How to perform the acceptable noise level (ANL) test with Equinox Evo

Click here to learn how to perform the acceptable noise level (ANL) test with the Equinox Evo clinical audiometer.

Interacoustics2026-10-02

How to perform the Alternate Binaural Loudness Balance (ABLB) test with Equinox Evo

Click here to learn how to perform the alternate binaural loudness balance (ABLB) test with the Equinox Evo clinical audiometer.

Interacoustics2026-10-02

How to use Ambient Noise with Equinox Evo

Click hear to learn how to use the Ambient Noise feature with the Equinox Evo clinical audiometer.

Interacoustics2026-10-02

How to perform basic audiometry with Equinox Evo

Click here to learn how to perform basic audiometry with the Equinox Evo clinical audiometer.

Interacoustics2026-10-02

Equinox Suite: How to set up your VRA

Learn how to set up your visual reinforcement audiometry (VRA) using the Equinox Suite software with the Equinox Evo clinical audiometer.

Interacoustics2026-10-02

Equinox Evo: How to set up the assistant monitor

Learn how to set up the assistant monitor functionality using the Equinox Suite software with the Equinox Evo clinical audiometer.

Interacoustics2026-10-02

Equinox Suite: How to use the VRA feature

Learn how to use the visual reinforcement audiometry (VRA) feature in the Equinox Suite software with the Equinox Evo clinical audiometer.

Interacoustics2026-10-02

How to perform Free Field (FF) Adjustment with Equinox Evo

Click here to learn how to perform free field adjustment with the Equinox Evo clinical audiometer.

Interacoustics2026-10-02

How to perform High Frequency Audiometry with Equinox Evo

Click here to learn how to perform high frequency audiometry with the Equinox Evo clinical audiometer.

Interacoustics2026-10-02

How to print with Equinox Evo

Click here to learn how to print with the Equinox Evo clinical audiometer.

Interacoustics2026-10-02

How to perform IA-AMTAS with Equinox Evo

Click here to learn how to perform the IA-AMTAS automated audiometry test with the Equinox Evo clinical audiometer.

Interacoustics2026-10-02

How to perform Masking Level Difference (MLD) with Equinox Evo

Learn how to perform the masking level difference (MLD) test with the Equinox Evo clinical audiometer.

Interacoustics2026-10-02

PC keyboard shortcuts with Equinox Evo

Click here to access the PC keyboard shortcuts for the Equinox Evo clinical audiometer.

Interacoustics2026-10-02

How to use pediatric noise with Equinox Evo

Learn how to use the pediatric noise stimulus with the Equinox Evo clinical audiometer.

Interacoustics2026-10-02

How to set up protocols in the Equinox Evo software

Click here to learn how to set up protocols in the software for the Equinox Evo clinical audiometer.

Interacoustics2026-10-02

How to perform pure tone audiometry with Equinox Evo

Click here to learn how to perform pure tone audiometry with the Equinox Evo clinical audiometer.

Interacoustics2026-10-02