How to perform the Stenger test with Equinox Evo

By Interacoustics

Required equipment

  • Equinox Evo
  • A patient response button
  • Calibrated headphones or insert phones

Test procedure for tone Stenger

Open the AC440 module.

Choose settings for Stenger using the input and output for channel 1 and channel 2 as illustrated below (Figure 1).

Figure 1: Man (Manual) and if preferred Multi Pulse should be selected for channel 1, and Sim (Simultaneous) for channel 2.

Typically, the instruction to the patient is not different than with normal pure tone audiometry. The patient does not need to be informed that stimuli are presented at both ears simultaneously.

Use channel 1 for the poorer ear and set the intensity level 10 dB below the claimed threshold.

Use channel 2 for presenting tones to the better ear and set the intensity level 10 dB above the threshold.

Present the tones to both ears simultaneously. Keep the intensity level of the better ear fixed and increase the intensity for the poorer ear in 5-dB steps.

If the hearing loss in the poorer ear is genuine, the patient will keep responding to the signal, as it is perceived in the better ear (negative Stenger). In case of a non-organic hearing loss, the patient will stop responding when the level exceeds the claimed threshold of the poorer ear (positive Stenger).

Test procedure for speech Stenger

Open the AC440 module and go to the speech screen.

Choose settings for Stenger using the input and output for channel 1 and channel 2 as illustrated below (Figure 2).

Figure 2: Rev (Reverse) and if preferred Multi Pulse should be selected for channel 1 and Rev (Reverse) or Sim (Simultaneous) for channel 2.

Clarify the test procedure for the patient. Explain that they will now hear speech, and the assignment is to repeat the presented words or numbers.

Use channel 1 for the poorer ear and set the intensity level 10 dB below the claimed threshold.

Use channel 2 for presenting speech to the better ear and set the intensity level to 10 dB above the threshold.

Present the speech material to both ears simultaneously. Keep the intensity level of the better ear fixed and increase the intensity for the worse ear in 5-dB steps.

If the hearing loss in the worse ear is genuine, the patient will keep repeating correctly to the speech signal presented to the better ear (negative Stenger). In case of a non-organic hearing loss, the patient will stop responding when the level exceeds the claimed threshold of the poorer ear (positive Stenger).

Presenter

Interacoustics

Available Training

How to perform the acceptable noise level (ANL) test with Equinox Evo

Click here to learn how to perform the acceptable noise level (ANL) test with the Equinox Evo clinical audiometer.

Interacoustics2026-10-02

How to perform the Alternate Binaural Loudness Balance (ABLB) test with Equinox Evo

Click here to learn how to perform the alternate binaural loudness balance (ABLB) test with the Equinox Evo clinical audiometer.

Interacoustics2026-10-02

How to use Ambient Noise with Equinox Evo

Click hear to learn how to use the Ambient Noise feature with the Equinox Evo clinical audiometer.

Interacoustics2026-10-02

How to perform audiometric masking with Equinox Evo

Learn how to perform audiometric masking with the Equinox Evo clinical audiometer.

Interacoustics2026-10-02

How to perform basic audiometry with Equinox Evo

Click here to learn how to perform basic audiometry with the Equinox Evo clinical audiometer.

Interacoustics2026-10-02

Equinox Suite: How to set up your VRA

Learn how to set up your visual reinforcement audiometry (VRA) using the Equinox Suite software with the Equinox Evo clinical audiometer.

Interacoustics2026-10-02

Equinox Evo: How to set up the assistant monitor

Learn how to set up the assistant monitor functionality using the Equinox Suite software with the Equinox Evo clinical audiometer.

Interacoustics2026-10-02

Equinox Suite: How to use the VRA feature

Learn how to use the visual reinforcement audiometry (VRA) feature in the Equinox Suite software with the Equinox Evo clinical audiometer.

Interacoustics2026-10-02

How to perform Free Field (FF) Adjustment with Equinox Evo

Click here to learn how to perform free field adjustment with the Equinox Evo clinical audiometer.

Interacoustics2026-10-02

How to perform High Frequency Audiometry with Equinox Evo

Click here to learn how to perform high frequency audiometry with the Equinox Evo clinical audiometer.

Interacoustics2026-10-02

How to print with Equinox Evo

Click here to learn how to print with the Equinox Evo clinical audiometer.

Interacoustics2026-10-02

How to perform IA-AMTAS with Equinox Evo

Click here to learn how to perform the IA-AMTAS automated audiometry test with the Equinox Evo clinical audiometer.

Interacoustics2026-10-02

How to perform Masking Level Difference (MLD) with Equinox Evo

Learn how to perform the masking level difference (MLD) test with the Equinox Evo clinical audiometer.

Interacoustics2026-10-02

PC keyboard shortcuts with Equinox Evo

Click here to access the PC keyboard shortcuts for the Equinox Evo clinical audiometer.

Interacoustics2026-10-02

How to use pediatric noise with Equinox Evo

Learn how to use the pediatric noise stimulus with the Equinox Evo clinical audiometer.

Interacoustics2026-10-02

How to set up protocols in the Equinox Evo software

Click here to learn how to set up protocols in the software for the Equinox Evo clinical audiometer.

Interacoustics2026-10-02