How to perform FM transparency verification

By Interacoustics

Note: FM transparency assumes that the hearing aid has already been programmed to match the target, according to the patient's hearing loss.

Note: The examples shown below are using the Affinity Compact system. However, the steps are the same.

Launch Affinity/Callisto Suite and select the REM module. Select the FM Transparency protocol from the drop-down menu (Figure 1).

Figure 1: FM Transparency protocol.

When the protocol has been selected, a setup guide will pop up with instructions on how to perform FM transparency verification (Figure 2). See below steps for further instructions.

Figure 2: FM Transparency setup guide. Note: this setup guide shows a different FM transmitter.

Step 1: Hearing aid alone (65 dB)

1. Select the 'HA Alone' test.

2. Attach the hearing aid to the coupler/coupler microphone.

3. Position the hearing aid in the test chamber (Figure 3).

Figure 3: Hearing aid placed in test chamber.

4. Click 'START' to measure the output of the hearing aid (Figure 4).

Figure 4: HA Alone measurement.

Step 2: Hearing aid and receiver (65 dB)

1. Leaving the hearing aid attached to the coupler, attach the FM receiver to the hearing aid OR connect the FM transmitter to the hearing aid with Bluetooth (depends on the type of FM transmitter) (Figure 5).

Figure 5: Setup for hearing aid and receiver measurement.

2. Make sure the FM transmitter is turned on but muted.

3. Click 'START' to measure the output of the hearing aid (Figure 6).

Figure 6: Hearing aid and receiver measurement. Note: The two curves should be very similar to each other at this step.

Step 3: FM microphone and the hearing aid

1. Place the hearing aids, still attached to the coupler microphone, outside of the test box.

2. Unmute the FM transmitter.

3. Position the FM transmitter inside the test box, facing down on the integrated speaker.

4. Place the reference microphone at the top of the FM transmitter. See Figure 7.

Figure 7: Setup for FM microphone and hearing aid measurement.

5. Click 'START' to measure the output of the hearing aid when input is from the FM microphone (Figure 8).

Figure 8: Setup for FM microphone and hearing aid measurement.

If you already now notice a significant difference between the curve from step 2 and the curve from step 3, you can presume that there is a problem with the FM microphone and that there is not transparency. Ensure that the test is conducted in a quiet environment, since the hearing aid is outside of the test box and the hearing aid microphones are still active.

Step 4: Calculate the average difference

1. Click on the table mode icon (Figure 9) to activate table mode (Figure 10).

Figure 9: Table mode icon.

Figure 10: Table mode.

2. Once in table mode, calculate a 3-frequency average of the differences between the hearing aid response in step 2 and the FM transmitter response in step 3 for the following frequencies: 750 Hz, 1000 Hz, and 2000 Hz (Figure 11). See example below.

Figure 11: Calculating a 3-frequency average.

If the average is 2 dB or less, FM transparency is achieved. If the average is more than 2 dB, adjust the FM.

Example of calculating the transparency

Using the numbers from Figure 11, let's calculate the transparency (Table 1).

Hearing aid + receiver FM mic + hearing aid Difference
750 Hz 74 72 2
1000 Hz 71 71 0
2000 Hz 77 75 2
Average 1,3*

Table 1: Example of calculating the transparency.

*Average difference is less than 2 dB. Therefore, transparency has been achieved.

Presenter

Interacoustics

Available Training

Acceptable Noise Level (ANL) Test

Learn how to perform the acceptable noise level (ANL) test with the Affinity 2.0 device.

Interacoustics2026-10-02

AC440 Protocol Setup

There are many different demands and wishes regarding how an audiometer should both function and appear depending on the specific situation and national standards. One of the greatest advantages of the AC440 module is the flexibility, enabling you to tailor the system according to your specific preferences in the so-called “test protocols”. This article describes basic functionality on how to work with the AC440 Protocols.

Interacoustics2026-10-02

Adult Hearing Aid Verification

Learn how to verifiy hearing aids in adults with the Affinity 2.0 device, using real ear measurements (REM).

Interacoustics2026-10-02

Basic Audiometry

Learn how to do basic audiometry with the Affinity 2.0 device.

Interacoustics2026-10-02

Binaural Real Ear Measurements

Learn how to perform binaural real ear measurements (REM) with the Affinity 2.0.

Interacoustics2026-10-02

BKB-SIN speech-in-noise test

In this quick guide, you will learn how to use the BKB-SIN speech-in-noise test provided with the Interacoustics Affinity 2.0 audiometer. Read more.

Interacoustics2026-10-02

How to perform CROS and BiCROS fitting

Learn how to fit CROS and BiCROS hearing aids with the Affinity 2.0 device.

Interacoustics2026-10-02

How to customize icons

Learn how to customize icons with the Affinity.

Dennis Mistry2026-10-02

How to perform frequency lowering verification

In this Affinity 2.0 quick guide, you'll learn how to determine if you should apply frequency lowering and how to verify the frequency lowering feature.

Interacoustics2026-10-02

Delta Values

Learn how to get a more accurate target match when fitting hearing aids with the Affinity.

Dennis Mistry2026-10-02

Hearing Instrument Testing

Learn how to perform hearing instrument testing with the Affinity 2.0 device.

Interacoustics2026-10-02

HIT440 Test Protocol

Learn how to set up protocols in the Affinity's hearing instrument testing module.

Interacoustics2026-10-02

How to Create a New Protocol in REM440

Learn how to create protocols in the Affinity's real-ear measurement module.

Leigh Martin2026-10-02

How to do an RECD Using the SPL60 Probe

Learn how to do an RECD using the SPL60 probe.

Leigh Martin2026-10-02

How to do Pure Tone Audiometry

Learn how to do pure tone audiometry with the Affinity.

Dennis Mistry2026-10-02

How to do Real Ear Measurement in Open Fit Hearing Aids

Learn how to do real-ear measurements in open-fit hearing aids.

Leigh Martin2026-10-02